OXITEST PLUS 7 OPERATING MANUAL
3.2
Powering Up, and Modifying the Oxitest Plus 7 Menu
To turn on the
Oxitest Plus 7
rotate the test probe, which is visible from the right hand side of the
Oxitest Plus 7
, 90 degrees clockwise from its storage compartment.
Oxitest Plus 7
will go
through a power-on self-test (POST). The LCD display will appear as shown below.
**Oxitest PLUS 7**
***Version: X.X***
After the POST is completed, the display will show the current test parameters as follows.
N-200 NO SIGNAL
97% 60 BPM 100%
When initially delivered, the default test parameters are set to Nellcor N-200, 97% SpO
2
, 60 BPM
and 100% pulse amplitude. If the parameters are changed, the
Oxitest Plus 7
will retain the
settings in memory and power up with those parameters on next use.
When received from the factory, all pulse oximeters with which
Oxitest Plus 7
is compatible will
be
ENA
bled in the
Pulse Ox
menu list. Not all oximeters will be pertinent to every user. To
reduce the number of oximeters presented in the menu list, individual oximeters can be
DIS
abled
with the
Setup Function
described below.
To enter the
Setup Function
, depress the
SpO
2
button while turning on the power to the
Oxitest
Plus 7
. The display shown below will appear for about 2 seconds.
**PULSE OX SETUP**
***Version: X.X***
The pulse oximeter information will then be displayed as shown below. The oximeter information
includes the manufacturer name, model number(s) and the type of sensor used (manufacturers may
use more than one type of sensor, each of which may require a different SpO
2
curve, or “R-Curve”).
Nellcor N-200,N-100C
SENSOR: Nellcor >ENA
Operation/Chapter 3
#
Page 11