OXITEST PLUS 7 OPERATING MANUAL
Datex and Invivo oximeters:
±1% at 97, 93% SpO
2
±2% at 90, 80, 70% SpO
2
1.3
Heart Rate
Variable from 20 to 250 beats per minute (BPM) in 1 BPM steps
Presets:
30, 60, 90, 120, 180, 240 BPM
Accuracy:
± 1 BPM
1.4
Pulse Amplitude
Variable from zero (no blood flow) to 100% (normal adult pulse) in 1% steps.
Presets:
100%, 30%, 10%, and 5%
Accuracy:
± 1%
1.5
Signal Artifact
Four preset simulations:
Movement
Tapping (Spike artifact)
Shivering (Tremor artifact)
Shake Table (2.5Hz Sinewave)
1.6
Auto Presets
Nine preset patient simulations. Factory default Auto Presets are:
Normal Adult
Hypoxia
Tachycardia
Bradycardia
Low Perfusion
No Perfusion
Movement Artifact
Neonate
Tremor (Shivering Artifact)
Auto Preset parameters may be user-programmed via the RS-232 port.
Specifications/Chapter 1
#