OXITEST PLUS 7 OPERATING MANUAL
3.5
Selecting an SpO
2
level
Oxitest Plus 7
simulates five (5) preset levels of oxygen saturation - 97, 93, 90, 80 and 70%.
Additional preset levels of 55% and 35% may also be provided, according to pulse oximeter
manufacturer, model number, and sensor (see Section 1.1).
The
SpO
2
preset is changed by pressing the
SpO
2
key on the front panel. The selected
SpO
2
preset
will be indicated on the standard operating screen at the lower left corner of the display, as follows.
N-200 NO SIGNAL
97%
60 BPM 100%
The
SpO
2
preset on power up will be the one retained from the last test performed.
The oxygen saturation simulation
can be fine-tuned in steps of 1% by means of the "ADJUST
SPO2" menu. To access this menu, press the
SpO
2
and
Pulse Ox
keys simultaneously. The display
will then change to the following:
ADJUST SPO2: 97%
Use the
•
and
–
keys associated with the
Heart Rate
key or the
Pulse Amp
key to adjust the
oxygen saturation as desired. Continuous adjustment is possible in 1% steps from 100% down to the
lowest SpO
2
preset provided, which may be either 70% , 55% or 35%, depending on the oximeter
model and/or sensor, as described above. To exit the "ADJUST SPO2" menu and return to the
standard operating screen, press either the
SpO
2
key or the
Pulse Ox
key.
Operation/Chapter 3
#
Page 14