Copyright © 2008
–
2019 DARE!! International
Page 10
2.4
Fiber Optic Cables
Use an extension fiber to connect the sensors to the plug-in card mounted in the Radi
C
entre
®
.
This extension fiber is a robust duplex fiber cable and uses dissimilar connectors to avoid
misconnections. The fiber optic cable with FC connectors feeds LASER light to the field sensor.
The fiber optic cable with FSMA connectors sends data from the sensor to the readout unit. Refer
to chapter 2.8 for fiber handling and maintenance.
To ensure safe and correct operation of the sensor, only use the Radi
S
ense
®
with the original
fibers. Do not use other fibers than those supplied by DARE!! Instruments. Fiber extension cables
are available on request. Contact your local reseller for more detailed information.
2.5
Radi
S
upply
®
Plug-In Card, Rear Panel
The following connections and indicators are found at the back panel of the Radi
S
upply
®
plug-in
card:
POWER ON
This green LED lights up when the power is on and the interlock is
closed.
LASER APERTURE
Connect to the E-field sensor through the fiber extension cable
with FC connectors.
LASER ON
This red LED lights up when the LASER is active. For safety reasons,
never disconnect the fiber optic cables when the LASER is switched
on.
OPTICAL DATA INPUT
Connect to the E-field sensor through the fiber extension cable
with FSMA connectors.
Figure 1: Radi
S
upply
®
plug-in card with
beam shutter
LASER APERTURE
OPTICAL DATA
INPUT
BEAM SHUTTER
SLIDE
‘POWER ON’ LED (GREEN)
‘LASER ON’ LED
(RED)