Copyright © 2008
–
2019 DARE!! International
Page 14
•
Do not activate the system if the fiber optic cables show any sign of damage or
tampering.
•
The Radi
C
entre
®
system is equipped with a remote interlock system. This
interlock system prevents inadvertent LASER radiation. For example, to prevent
LASER radiation when someone enters a shielded room and steps on a fiber.
•
The remote interlock connection should be connected to an ‘emergency mas
ter
disconnect’ and in series with the room door or fixture interlocks.
•
A visual ‘LASER ON’ indicator LED will light up when the LASER is activated. This
LED on the front panel of the Radi
C
entre
®
system serves as a reminder to the
operator that one or more LASERS are switched on.
Radi
C
entre
®
2-slot and 7-slot specific:
•
A ‘LASER Code’ is required to activate the
Radi
C
entre
®
2-slot and 7-slot systems.
This LASER code enables the power supply to all installed (LASER) modules.
•
To prevent accidental activation of the LASER, an ‘Acknowledge’ button will appear
directly after the LASER ‘Start’ button is pressed. The LASER will on
ly be activated if
this button is pressed within the ‘4 seconds timeframe’.
Radi
C
entre
®
single-slot specific:
•
To start the Radi
C
entre
®
1-
slot system (which has no touchscreen display), a ‘Start’
button needs to be pressed for at least 3 seconds to activate the LASER. If you are
interrupted during the activation process (and release the ‘Start’ button by
accident), the LASER will not be activated.
•
To prevent accidental activation of the LASER, an auditory warning will alert you of
the LASER activation procedure (if the
‘Start’ button is being pressed). To interrupt
the activation process, all you must
do is release the ‘Start’ button.