5.2
Commands
5.2.1
General commands for the Radi
S
ense
®
Instruction
Command
Reply
Set range
3
" Ra\r "
Where
a
is the range:
1
: 0 to 10V/m
2
: 10 to 30V/m
3
: 30 to 100V/m
4
: 100V/m to end of spec.
" :Rx "
Where
x
is the range.
Axis on/off
" Axyz\r "
Where
x, y
and
z
are:
D
for Disable
E
for Enable
" :A "
Get field
" Da\r "
Where
a
is notation:
1
: short notation
2
: long notation
3
: all axis, short notation
" :Dxx.xxuuu "
" :Dxx.xxuuurrrobxyz "
" :Dxx.xx;yy.yy;zz.zzuuu "
Where:
xx.xx
(or
y
or
z
) is the reading.
uuu
is the unit:
_V_
for V/m
mW2
for mW/cm²
_V2
for (V/m)²
rrr
is the recorder out value
from 0 to 255.
o
is over range indicator:
N
is Ok
O
is over range
b
is battery status:
N
is Ok
W
is warning
F
is fail
xyz
is the axis on/off status:
D
for Disable
E
for Enable
*
This table continues on the next page.
3
This command does not affect the actual internal Radi
S
ense
®
measuring range, but is included for ‘Holaday
®
compatible’. However, the set range value is used to calculate the recorder values in the ‘Get field’ command.