To connect the Radi
S
ense
®
sensor, the slide in the beam shutter must be opened. Use a small
tool (for example a tweezer) to open the slide in the beam shutter. Hold the slide open and insert
the FC connector at the same time. While inserting the FC connector, keep the notch facing to the
right, to mate with the slot in the socket.
Figure 3: Opening the slide (left) and inserting the FC connector with the notch facing right (right)
When the FC connector is inserted in the laser aperture,
plug-in the FSMA connector in the optical data input.
Do not use any tools or excessive force to tighten the optical
connectors!
If both connectors are properly connected, the Radi
C
entre
®
can be switched on and the LASER can be activated to start
measurements.
Be sure that the FC connector is lined correctly (Figure 4a)
otherwise the LASER will not start correctly.