5.
Connect the plug-in card to the desired device(s).
Place the Radi
S
ense
®
E-field sensor where the field strength is to be measured.
Clean the ends of the fibers and connect the cables to both the sensor and the
plug-in card. Make sure the latching pin of the FC connector fits correctly in the
slot of the chassis connector (in line coupling).
Figure 13:
Slide the ‘notch’ correctly into the ‘slot’ to connect the
FC connectors
6.
Connect the Radi
C
entre
®
system to your PC using any of the available interfaces
of the Radi
C
entre
®
system; USB, RS232, Ethernet or IEEE-488 (optional).
7.
Place the interlock plug of the plug-in card
into the interlock connector of the
Radi
C
entre
®
.
The hardware installation for the plug-in card is now complete. The user can control the plug-in
card either through the touchscreen on the Radi
C
entre
®
system (only available for the 2 and 7-
slot versions), or by using the control commands in combination with an external software
package such as the Radi
M
ation
®
EMC test software.