Figure 25: Line scan, Fixed Frame, Level Trigger (Roll-Over)
CVI File (VIC) Parameters Used
The VIC parameters listed below provide the control functionality for virtual frame reset. Sapera
applications load pre-configured CVI files or change VIC parameters during runtime.
Note: Sapera camera file parameters are easily modified by using the CamExpert program.
External Frame Trigger Enable = X, where: (with Virtual Frame Trigger enabled)
•
If X = 1, External Frame Trigger is enabled
•
If X = 0, External Frame Trigger is disabled
External Frame Trigger Detection = Y, where:
(with Virtual Frame Trigger edge select)
•
If Y= 1, External Frame Trigger is active low
•
If Y= 2, External Frame Trigger is active high
•
If Y= 4, External Frame Trigger is active on rising edge
•
If Y= 8, External Frame Trigger is active on falling edge
•
If Y= 32, External Frame Trigger is dual-input rising edge
•
If Y= 64, External Frame Trigger is dual-input falling edge
Note:. For dual-input triggers, Trigger Input #1 signals the start of the frame trigger, Trigger
Input #2 signals the end of the frame trigger.
External Frame Trigger Level = Z, where:
(with Virtual Frame Trigger signal type)
•
If Z = 1, External Frame Trigger is a TTL signal
•
If Z = 2, External Frame Trigger is differential signal (RS-422)
•
If Z = 8, External Frame Trigger is a 24V signal
•
If Z = 64, External Frame Trigger is a 12V signal
Note: For information on camera configuration files, see the Sapera Acquisition Parameters
Reference Manual (OC-SAPM-APR00).