The information contained in this document is not subject to Export Control Regulation in Canada.
Calibir GXM640 IR Camera User's Manual
Operational Reference
89
Display Name
Feature & Values
Description
View
Temperature
Temperature FPN will be automatically
recalibrated if the camera’s
internal temperature drifts more
than a certain threshold since
the last FPN calibration. The
threshold is configured using
autoShutterMaxTempDiff.
Time
Time FPN will be automatically
recalibrated if a certain time
(configured by
autoShutterMaxDelay) has
elapsed since last FPN
recalibration.
TimeTemperature
TimeTemperature FPN recalibration will be
triggered if either the
temperature or time criteria is
exceeded.
Auto Shutter Min Delay
(in s)
autoShutterMinDelay
Sets the minimum time between
two automatic FPN recalibrations.
Beginner
DFNC
Auto Shutter Max Delay
(in s)
autoShutterMaxDelay
Sets the maximum time between
automatic FPN recalibrations.
Used when autoShutterMode is
either Time or TemperatureTime.
Beginner
DFNC
Auto Shutter Max
Temperature (in
Celsius)
autoShutterMaxTempDiff
Sets the camera temperature
difference, in degrees Celsius,
that will trigger automatic FPN
recalibration.
Beginner
DFNC
Shutter State
shutterState Specifies the current shutter
state.
Beginner
DFNC
Open
Open Shutter is open.
Closed
Closed Shutter is closed.
Shutter Activation
Count
shutterActivationCount
Number of times the internal
mechanical shutter has been
activated. This value is saved in
camera memory and persists
when the camera is reset or
rebooted.
Guru
DFNC
FPN Target Mode
shutterTarget
Selects the shutter target.
Expert
DFNC
External Shutter
External The shutter is an external shutter
outside the camera body.
Internal Shutter
Internal Internal camera mechanical shutter.
Internal Advanced
InternalAdvanced FPN calibration will use the internal
mechanical shutter, but with
advanced modelling which provides
more uniform images. This option
may not be available for cameras
that were shipped prior to release
2.41.