The information contained in this document is not subject to Export Control Regulation in Canada.
112
Operational Reference
Calibir GXM640 IR Camera User's Manual
Display Name
Feature & Values
Description
View
PTP Clock Offset
Adjust Factor
ptpClockOffsetAdjustFactor
Sets the PTP clock adjust factor. When the camera
is a PTP slave, it adjusts its clock to match the
PTP master's clock. When the camera receives a
PTP Sync message, it computes the clock offset
between the camera and the PTP master. This
offset value is multiplied by
ptpClockOffsetAdjustFactor and used to adjust the
camera clock offset, compensating for past errors.
When ptpClockOffsetAdjustFactor is 1.0, the
camera instantly adjusts to the master clock,
making it more sensitive to variations in network
transmission time of PTP Sync packets.
Guru
DFNC
PTP Clock Offset Max
Adjust (in ns)
ptpClockOffsetMaxAdjust
Maximum clock offset adjustment, in
nanoseconds, that can be done when the camera
receives a PTP Sync (and PTP Servo Status is
“Locked”).
Guru
DFNC
PTP Clock Speed
Adjust Factor
ptpClockSpeedAdjustFactor
Sets the PTP clock speed adjust factor. When the
camera is a PTP slave, it adjusts its clock to match
the PTP master's clock. When the camera receives
a PTP Sync message, it computes the clock offset
between the camera and the PTP master. This
offset value is multiplied by
ptpClockSpeedAdjustFactor and used to adjust the
camera clock offset, compensating for future
errors. When ptpClockSpeedAdjustFactor is 1.0,
the camera instantly adjusts to the master clock,
making it more sensitive to variations in network
transmission time of PTP Sync packets.
Guru
DFNC
PTP Clock Speed Max
Adjust (in ns per s)
ptpClockSpeedMaxAdjust
Maximum clock speed adjustment, in
nanoseconds per second, that can be done when
the camera receives a PTP Sync.
Guru
DFNC
Alarm Selector
alarmSelector
Specifies the alarm to configure.
Beginner
DFNC
Alarm 1
Alarm1 Selects Alarm1.
Alarm 2
Alarm2 Selects Alarm2.
Alarm 3
Alarm3 Selects Alarm3.
Alarm Source
alarmSource Selects which value will be monitored for this
alarm.
Beginner
DFNC
Off
Off Deactivates this alarm.
Zone 1 Average
Zone1Average Alarm will be triggered if the average pixel value
in Zone 1 meets a certain criteria.
Zone 1 Pixel Count
Zone1PixelCount Alarm will be triggered if a minimum number of
raw sensor pixels in Zone 1 meet a certain
criteria.
Zone 2 Average
Zone2Average Alarm will be triggered if the average pixel value
in Zone 2 meets a certain criteria.
Zone 2 Pixel Count
Zone2PixelCount Alarm will be triggered if a minimum number of
raw sensor pixels in Zone 2 meet a certain
criteria.
Camera Temperature
CameraTemperature Alarm will be triggered if the camera's internal
temperature (measured as the Processor's
temperature) exceed a certain criteria.
Alarm Source Value
alarmSourceValue
Returns the latest value of the source being
monitored by the selected Alarm.
Beginner
DFNC
Criteria Type
alarmCriteria
Selects which criteria must be met by the
monitored value before an alarm is triggered.
Beginner
DFNC
Less Than
LessThan For Pixel Count-type alarms, a pixel will be
counted if its value is less than the Criteria
Threshold. For other alarms, alarm will be
triggered if the Alarm Source Value is less than
the Criteria Threshold.