The information contained in this document is not subject to Export Control Regulation in Canada.
132
Technical Specifications
Calibir GXM640 IR Camera User's Manual
External Output AC Timing Characteristics
The graphic below defines the test conditions used to measure the Calibir external output
AC characteristics, as detailed in the table that follows.
t
t
Output Control Signal
Output
100%
90%
10%
t
d1
t
rise
t
fall
t
d2
Output
Output Common Power
R
Load
Control
Signal
Opto-coupled Output (Pin 6):
AC Characteristics
Note: All measurements subject to some rounding.
Output
Common
Power
Output
Current
Output
Voltage
t
d1 (
µs)
Leading Delay
t
rise (
µs)
Rise Time
t
d2 (
µs)
Trailing Delay
t
fall (
µs)
Fall Time
3V
8 mA
1.55V
1.5us
8.96us
4.5us
10us
16 mA
1.25V
2.1us
8.58us
8.3us
8.6us
5V
8 mA
3.32V
1.6us
4.6us
12.6us
10.7us
16 mA
3.1V
1.7us
6.3us
10.0us
8.5us
21 mA
2.9V
1.8us
7.2us
8.8us
7.8us
12V
8 mA
10.2V
1.9us
2.8us
15.9us
13.6us
16 mA
10.0V
2.0us
3.9us
11.0us
9.9us
24 mA
9.5V
1.9us
3.0us
9.7us
6.3us
24V
8 mA
22.5V
2.2us
3.0us
18.5us
13.6us
16 mA
21.65V
2.2us
2.6us
12.8us
11.0us
24 mA
21.3V
2.2us
2.7us
11.4us
8.1us