DFX-7
Ultrasonic Flaw Detector
55
6) Press the
OK
key to set the
X-OFFSET
and return to the menu screen, or
ESC
to cancel entering the
X-OFFSET
.
Note:
The
X-Offset
is used for purposes of marking, and is physically
measured with a ruler from the index point to the front edge of the transducer.
Curved Surface Outside Diameter
1) Press the
UP
and
DOWN
arrow keys until
CSC OD
is highlighted, followed
by pressing the
LEFT
and
RIGHT
arrow keys to scroll the value. When the
correct
CSC OD
is displayed, proceed to the next setting.
2)
Alternatively
, press the
ENTER
key to display the
Digits Edit Box
.
3) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
4) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
5) Repeat steps
3 & 4
until the
CSC OD
is correctly displayed.
6) Press the
OK
key to set the
CSC OD
and return to the menu screen, or
ESC
to cancel entering the
CSC OD
.
Note:
The
CSC OD
value is only used for curved surfaces of longitudinally
welded pipes.
Show Legs (Skips)
1) Press the
UP
and
DOWN
arrow keys until
SHOW LEGS
is highlighted,
followed by pressing the
LEFT
and
RIGHT
arrow keys to toggle on/off.
Note:
When
SHOW LEGS
is enabled the background of the A-Scan area will
display strips with alternating color for an easy reference of the leg.
.