46
Adjusting the Gate to Calibrated Range
1) Press the
MEAS
key once to activate the measure menu items. Press the
MEAS
key multiple times to move right and the
ESC
key multiple times to
move left, until the
GATE1
cell is highlighted.
2) Press the
LEFT
and
RIGHT
arrow keys to move the gate start to a position
less than 1 inch.
3) Press the
MEAS
key once to highlight the
WIDTH1
cell.
4) Press the
LEFT
and
RIGHT
arrow keys to move the gate start to a position
greater than 4 inches.