Ultrasonic Flaw Detector
13
ANGLE:
Used to set the angle of the transducer, and used in the calculations above.
Refer to page 51 for further info.
X-OFFSET:
This feature is used to calculate the surface distance in the trigonometry
function, as is a physical measurement from the index point of the probe to the front
of the wedge/can/shoe/case. Refer to page 51 for further info.
CSC OD:
Curved surface correction to compensate for changing sound path
distance and angle. This feature is used in conjunction with ‘thickness’ in order to
calculate a correction factor. Refer to page 51 for further info.
SHOW LEGS:
When this feature is enabled, the Legs or Skips will be displayed in
the A-Scan area with alternating background stripped areas for easy reference.
Refer to page 55 for further info.
3.9 AWS – Menu
AWS:
This feature toggles the AWS feature on/off. Used in the analysis of welds
according to American Welding Society’s structural welding code D1.1-94, to
calculate the Indication Rating (IR). Refer to page 56 for further info.
REFERENCE GAIN:
The gain (in dB) required to set the indication level to a specific
percentage of full screen height (FSH). This value will typically be set somewhere
between 20%–90%. Refer to page 56 for further info.
SHOW CURVE:
Toggles the AWS correction curve on/off, and the curve is
displayed on the screen. Refer to page 56 for further info.
% SCRN HEIGHT:
This feature enables the user to set the reference level of the
TCG as a percentage of full screen height (FSH). Refer to page 56 for further info.
3.10 TCG – Menu
TCG:
This feature toggles the TCG feature on/off. The Time Corrected Gain (TCG)
feature displays reflectors of equal size at equal A-Scan amplitudes, regardless of the
reflector’s depth in the test material. Refer to page 67 for further info.
DRAW:
Toggles the DRAW mode on/off, and used to draw the TCG curve. Refer to
page 67 for further info.
SHOW CURVE:
Toggles the TCG correction curve on/off, and the curve is displayed
on the screen. Refer to page 67 for further info.
% SCRN HEIGHT:
This feature enables the user to set the reference level of the
TCG as a percentage of full screen height (FSH). Refer to page 67 for further info.