3. Wait for the AMS readings to stabilize and then click
Next
to take the first calibration reading.
After a few seconds, the wizard displays the Step 3 of 4 Second Calibration Reading
Dialog.
W
AFER
S
EN
Rotate the AMS 180 degrees and align the AMS with the four marks you made in the earlier steps, as
shown below. Be sure to line up all four marks, so that the rotated AMS is as close as possible to the same
position it was in when you took the first calibration reading.