115
2. Allow the Pitch and Roll values to stabilize (around 30 seconds) and then hit
Next
.
This step calls for marking the position of the AMS on the reference surface. For 300mm or 200mm wafers
with notch, place a mark on the surface (or on tape on the surface) next to the edge of the AMS at the
notch. Place three additional marks on the surface next to the edge of the AMS at the scribe marks that
are 90 degrees, 180 degrees, and 270 degrees from the notch or flat side. Always mark all four positions.
Содержание AMS-150C
Страница 129: ...129 Dimensional Drawings Here are the dimensional drawings for the AMS AMS 150C Dimensions millimeters ...
Страница 130: ...130 AMS 200mm Dimensions millimeters ...
Страница 131: ...131 AMS 300C Dimensions millimeters ...
Страница 132: ...132 AMSR Dimensions millimeters ...
Страница 137: ...137 Appendix B Psychrometric Chart ...
Страница 143: ...143 Figure A 2 AHS Mode MultiReview Main Screen ...