27
Monitoring the Operating Temperature of the AMS
The operating range for the AMS is 20
–
70 °C. The AMS can withstand exposure to higher temperatures for very
brief periods, if not in direct contact with a heating element. The
Temperature
monitor in the
AMS wafer status
of
the MultiView window shows the current operating temperature of the AMS using a numeric readout and a bar
graphic (as shown below).
Low
Optimum
Warning
Danger
<20 °C
20
–
70 °C
71
–
80 °C
> 80 °C
The temperature bar changes color to indicate where the current temperature is relative to the accurate operating
range, using the following color-coded bars.
•
Blue
Less than 20 °C. The AMS is operating below the range, where it produces accurate readings.
•
Green
20
–
70 °C. The AMS is operating in its optimum temperature range, where it produces readings meeting the
specified accuracy.
•
Orange
71
–
80 °C. The AMS is operating above the range, where it produces the most accurate readings, but not so hot
that the AMS will be damaged.
•
Red
Greater than 80 °C. The AMS is operating at a temperature so high that it might be damaged.