73
If the self-test fails during operation and defibrillation cannot be administered, the device will
instruct you to get a replacement defibrillator and will guide CPR by voice. To learn more about
the error, first press the Power Button to turn off the device. Press and hold down the i-Button,
and the device will notify the error by voice and LCD Screen, and then automatically shut down.
For further details, please refer to [Section 7.3: Troubleshooting] of these Instructions for Use.
Since the CU-SP2 performs a self-test on a daily basis, you do not need to frequently perform
a self-test for the battery pack. Frequently self-testing the battery pack consumes battery
power and shortens the battery life.
Periodically check if the i-Button flashes in red in order to prepare for emergencies.
If the i-Button flashes in red, please refer to [Section 7.3: Troubleshooting] in these
Instructions for Use.