44
5. After Using the i-PAD CU-SP2
5.1 Maintenance After Each Use
Check the device for signs of damage and contamination. If there is any damage or
contamination, please refer to [Section 6.2.3: Cleaning the i-PAD CU-SP2] in these Instructions
for Use.
Conduct a self-test on the battery by referring to [Section 7.1: Self-Tests] in these Instructions
for Use. If the device shuts down normally after running a self-test on the battery, the device
status is normal. The i-PAD CU-SP2 uses disposable pads. Dispose of the used pads and
replace them with new pads after checking their expiration date. For further details on replacing
the pads, please refer to [Section 6.2.2: Replacing the Pads] in these Instructions for Use.
You should use only the defibrillator pads provided by the manufacturer.
Do not open the pad packaging until immediately before use. Since the adhesive material on
the disposable pads starts to dry out as soon as the package is opened, the pads will become
unusable after a certain amount of time has elapsed, regardless of the expiration date.