75
7.3 Troubleshooting
The device informs you of its current status or of problems via status indicators, beeps, and/or
voice instruction. Refer to the following for details:
Symptom/Voice Instruction
Cause
Resolution
Voice Prompt :
“Low battery”,
“Replace the battery with a new
one.”
The battery is low.
Replace the battery with a
new one.
Voice Prompt :
“Plug the pads connector into the
device.
”
The Pads Connector is
disconnected
Ensure the Pads
Connector is properly
connected.
Voice Prompt :
” Used pads”,
“Replace the pads with a new one”
The pads have been
previously used.
Replace the pads with a
new one.
Voice Prompt :
” The pads are beyond their
expiration date”,
“Replace the pads with a new one”
The pads have expired.
Replace the pads with a
new one.
Voice Prompt :
” Press the pads firmly to the bare
skin of the patient
”
The pads are not properly
attached to the patient's
skin.
Check if the pads are
securely attached to the
patient's skin.
Voice Prompt :
” No shock delivered”
The pads are not properly
adhering to the patient's
skin.
Press the pads firmly to the
patient's skin.
Shave chest hair or wipe
off moisture if necessary
before attaching the pads.
Voice Prompt :
” Shock button was not pressed”
Although an electric shock
is needed, the Shock
Button was not pressed
within 15 seconds.
Deliver an electric shock by
pressing the Shock Button
with the next voice
instruction.
If the problem cannot be solved during an emergency, you should follow the following steps:
①
Quickly replace the defibrillator if possible.
②
If no replacement device is available, check the patient's condition and perform CPR as
necessary. Continuously check the patient's condition and perform CPR until the
emergency medical services arrives.