28
Step 2: Press the Shock Button when instructed.
The device acquires and analyzes the patient's ECG immediately after being connected.
According to the device settings, automatic analysis will become available, along with ANALYZE
and CHARGE. If the device is set to automatic analysis, the device will automatically start
analyzing the ECG as soon as the pads are attached to the patient.
Do not touch the patient when the device instructs you not to touch the patient. The ECG
analysis may become inaccurate if you touch the patient during the analysis.
If the patient needs defibrillation after the ECG analysis, the device will do the following:
The device announces that a defibrillation shock is needed, and
instructs you to keep away from the patient.
When armed, the device will continuously beep while the Shock
Button flashes in orange.
The device instructs you to press the flashing orange Shock Button.
You should press the Shock Button at this time.
When the Shock Button is pressed, the device delivers a defibrillating shock to the patient. If
defibrillation is properly done, the device reports that an electric shock has been delivered.
After shock delivery, the device indicates that you may touch the patient and issues voice
instructions on CPR.
If the flashing Shock Button is not pressed within 15 seconds, the device will cancel the shock
delivery and disarm. Then the device issues CPR instructions.