If you want to stop charging, press the 'DISARM' button using Soft Key 3. When pressed, the
device will cease charging and dump the shock energy internally.
When armed, the Shock Button will flash in orange to signal readiness for defibrillation. At this
time, you can administer defibrillation by pressing the Shock Button.
If the flashing Shock Button is not pressed within 15 seconds, the device will automatically
cancel the shock delivery and disarm.
4.11.3 Using R-Sync
When the device is switched to manual mode
, Soft Key 2 will be activated as ‘SYNC ON’.
Pressing Soft Key 2 will display the SYNC symbol in the upper-center of the LCD Screen and
enable administration of R-Sync energy.
Using R-Sync will detect the R-wave of the patient's ECG, and display the R-Sync mark on the
LCD Screen with a short beep.
At this time, you may charge shock energy by pressing 'CHARGE' using Soft Key 3 if you think
that synchronous cardioversion is necessary.
Press the Shock Button to administer synchronous cardioversion. The device will automatically
administer synchronous cardioversion if R-wave is detected.
To stop using R-
Sync, press ‘SYNC OFF’ using Soft Key 2.