Section XV – Electromagnetic Conformity Declaration
The Kangaroo Joey enteral feed and flush pump with pole clamp has been built and tested according to UL60601-1,
EN60601-1:2006, CAN/CSA C22.2 No. 60601-1-08, and EN60601-1-2 Standards.
The Kangaroo Joey enteral feed and flush pump with pole clamp is intended for use in the electromagnetic environment
specified below. The user of the pump should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
(CISPR 11)
Group 1
The Kangaroo Joey enteral feed and flush pump with
pole clamp uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
(CISPR 11)
Class B
The Kangaroo Joey enteral feed and flush pump with pole
clamp is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
(IEC 61000-3-2:2005 + A1:2008 + A2:2009)
Class A
Voltage fluctuations/ flicker emissions
(IEC 61000-3-3:2008)
Complies
Radiated Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-3:2010)
Complies
Conducted Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-6:2003 +
A1:2004 + A2:2006)
Complies
Power Frequency Magnetic Field Immunity
(EN60601-1-2 / IEC 61000-4-8:2009)
Complies
Voltage dips and sags Immunity
(EN60601-1-2 / IEC 61000-4-11:2004)
Complies
Electrical Fast Transient / Bursts Immunity
(EN60601-1-2 / IEC 61000-4-4:2012)
Complies
Electrostatic Discharge Immunity
(EN60601-1-2 / IEC 61000-4-2:2008)
Complies
Surge Immunity
(EN60601-1-2 / IEC 61000-4-5:2005)
Complies