Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
10
Loading Pump Sets
The Kangaroo Joey enteral feed and flush pump with pole clamp will indicate
LOAD A SET
Figure 5A. LOAD A SET opening
menu screen for Continuous
Feeding mode (EZ Pump Mode
not activated).
in the screen’s status line if a pump set is not installed. A blinking icon of a pump set will also
appear in the upper left corner of the screen. Figure 5A shows the screen for continuous
feeding mode and Figure 5B shows the screen for intermittent feeding mode (intermittent
feeding mode not available in EZ Pump Mode).
To load a pump set, follow the steps below:
1. With the pump display facing you, open the blue transparent door that encloses the rotor
and pump set loading area.
2. Load the pump set by holding the finger tab on the valve, then inserting the valve into
the valve slot on the left. (Figure 6A)
3. Grasp the MISTIC retainer end (black ring retainer) and wrap the tubing counterclockwise
Figure 5B. LOAD A SET opening
menu screen for Intermittent
Feeding mode (EZ Pump Mode not
activated).
around the rotor.
Avoid overstretching the silicone tubing. (Figure 6B).
4. Carefully pull the MISTIC retainer end left and upwards to position it over the slot on the
left, and then lower the retainer into the slot. Push down on the retainer then release to
ensure proper engagement. (Figure 6B)
5. Push the valve’s finger tab to the right to ensure proper seating.
6. Close the blue transparent door, checking to make sure the valve finger tab is captured by
the slot in the blue door.
Note:
To avoid nuisance alarms, make sure the pump set exterior is free from droplets of
water or formula.
7.
The top of the starting volume of formula should be 6 inches above the pump (Figure 4B).
Figure 6B.
Set Loading Door
Figure 6A.
Pump Set
Valve
Power
Button
Finger Tab
MISTIC
Retainer
End
Valve Slot
RetainerSlot