Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
8
Section V — Instructions For Use
Quick Start
1. Press
POWER
on lower, right-hand corner of control panel.
2. Open blue door enclosing pump set loading area.
3. To load pump set (refer to Figure 4):
• Grasp finger tab on valve and insert into front pocket (1).
• Grasp black ring retainer and wrap tubing around rotor (2).
Pull up on retainer and insert into back pocket (3).
Avoid overstretching the silicone tube.
• Push finger tab to right to ensure that valve is fully seated in
the pocket.
• Close blue door.
(
Note:
display status line should read
SET LOADED
).
4.
Suspend the pump set bag so that the top of the
starting volume of formula is 6 inches above the
pump (Figure 4B).
5. To automatically prime the pump, press
PRIME PUMP
and
AUTO PRIME
.
For feed & flush sets, auto priming will prime both lines.
6. To directly control pump priming, press
PRIME PUMP
and then press and hold the
HOLD TO PRIME
button(s) . If a feed & flush set is used, remember to first press
HOLD TO PRIME FLUSH
until the flushing fluid has reached the valve, and then press
HOLD TO PRIME FEED
until the feeding fluid has reached past the valve and down to
the distal end connector.
7. Use buttons on left side of LCD to set the feeding parameters. Set flushing parameters
if a feed & flush set is loaded. The continuous or intermittent feed mode selection is made
under the
MORE
options menu.
8. Press
RUN
when ready to start. The screen will display
RUNNING
.
9. To stop, press
HOLD
or press and hold the
POWER
button.
General Startup
Placement/Mounting
There are two recommended placement methods for the Kangaroo Joey enteral feed and flush pump with pole clamp.
• Attached to a vertical IV pole via the pole clamp peripheral device, included with the pump.
• Placed on any stable surface.
A/C Power Operation
Plug the Kangaroo Joey enteral feed and flush pump with pole clamp into an A/C power outlet for normal operation. When the
pump is not plugged in, or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
Figure 4A. Installation of pump set. (See Figure 6 for detailed view.
Warning: Do Not Overstretch Tube.
1
3
2
Black Ring
Retainer
(MISTIC)
Finger
Tab
Valve
Rotor
182541
6 in
(15.2 cm)
6 in
(15.2 cm)
Figure 4B. Correct Fluid Level.