Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
28
Section X — Specifications and Symbols
Specifications
TUV Listing
Medical Electrical Equipment
Kangaroo Joey Enteral Feed and Flush Pump with Pole Clamp
(1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with
ES60601-1:2005, UL60601-1, IEC 60601-1:2005.
(2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance with
CAN/CSA C22.2 No. 60601-1-08.
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump Sets
Kangaroo Joey Feed-Only Set or Feed and Flush Set.
Feeding Formula Delivery Rate
1-400 mL/hr in 1 mL increments
Feeding Formula VTBD
1-3000 mL in 1 mL increments
Bolus Volume
1-3000 mL in 1 mL increments
Number of Boluses
1-99
Bolus Interval
1-24 hours in 1-hour increments
Flushing Solution Dose Range
10-500 mL in 1 mL increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
7% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 15.24 cm (6”) ± 0.76 cm (0.3”) above the top of the closed
pump door, at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a new Kangaroo Joey feeding set for no longer than the
recommended 24 hours of maximum usage. Confidence intervals for accuracy are based upon those included in the ANSI/AAMI
ID26-1992 American National Standard for Infusion Devices. Reported accuracy is based upon the time-based accuracy calculation in
this standard, as applicable for enteral feeding. Since feeding accuracy of the Kangaroo Joey enteral feed and flush pump with pole
clamp is based upon formula and not water, a 4% accuracy shift should be applied when testing the Kangaroo Joey enteral feed and
flush pump with pole clamp using water. (Water will appear to overdeliver by 4%.)
Occlusion Pressure
15 psi (103 kPa) Nominal