Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
27
Pump Set Dislodged Error
(Fixed Audible Volume)
Figure 30. Pump Set Dislodged
Error screen.
The
PUMP SET DISLODGED
screen will appear if the black ring retainer (MISTIC) is not properly
loaded in the MISTIC pocket in the pump set loading area. Check and correct the positioning of the
MISTIC retainer if possible. If the error cannot be resolved, load a new pump set, prime it, and press
CONTINUE
to restart the feeding, Figure 30.
If the error cannot be resolved, press
POWER
to stop operation of the pump and put a different
pump into service.
Battery Low
(Fixed Audible Volume)
Figure 31. Battery Low error
screen. Indicates that the battery
must be recharged immediately.
The
BATTERY LOW
screen appears and the alarm beeps continuously when the battery needs to be
recharged. There is approximately 15 minutes of battery life remaining when this screen appears,
Figure 31.
Plug the A/C power adapter in to begin recharging. The pump will automatically return to the
screen that was active prior to the error. The battery will charge continuously whenever the pump is
plugged into a wall outlet. 6 hours of charging is required to fully recharge the battery pack.
If this screen appears while the A/C adaptor is plugged in to the pump, check to make sure the A/C
adaptor plug is pushed all the way in so it is flush with the back of the pump.
Feeding Complete
(Adjustable Audible Volume)
Figure 32. Feeding Complete
notification screen.
The
FEEDING COMPLETE
information screen appears after completion of the programmed
feeding, Figure 32. Press
POWER DOWN
to turn off the pump. Press
CONTINUE
to return
to the
SET LOADED
opening menu, Figures 7A-7D.
Figure 33. Running screen with the >
24 HRS of Pump Set Usage Indicator in
upper left of the screen.
Set Use >24 Hours Warning
The pump set usage warning indicator will blink on the
RUNNING
screen if a pump set has been used for 24 or more
hours (hours actually running). It is recommended to replace
pump sets after this length of usage. This icon is only an informational message and does not
require action.
Figure 33 shows an example of the
RUNNING
screen, where a feed & flush set has been in use
for more than 24 hours. The informational warning indicator shows up in the upper left on
the screen.
LED Array
The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially
in darkened rooms.
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in
HOLDING
mode, or
one of the following errors:
HOLD ERROR
,
FEED ERROR
,
FLOW ERROR
, and
FLUSH ERROR
.
A red light status indicates a critical pump error. The screen will show one of the following error messages:
SYSTEM ERROR
,
PUMP
SET DISLODGED
, or
ROTOR ERROR
.