Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
14
Continuous Mode — Adjust Feed
Continuous Mode Feeding
For
CONTINUOUS
feeding mode, the
ADJUST FEED
option is available in the opening menus (Figure 7A for feed-only sets or Figure
7B for feed & flush sets). Select this option to program the feeding parameters from the
ADJUST FEED
screen, Figure 11A.
Press
FEED RATE
in Figure 11A to define the rate of delivery. Press the buttons in Figure 11B to program the rate from 1 to 400
in increments of 1 mL/hr. Press
ENTER
(top button) to exit the menu.
Press
FEED VTBD
in Figure 11A to define the Volume To Be Delivered. Press the buttons in Figure 11C to program the volume
from 1 to 3000 mL in increments of 1 mL. Press
ENTER
(top button) to exit the menu. If the
FEED VTBD
is set to zero, the pump
will run until the supply is exhausted.
NOTE:
When a
FEED VTBD
is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to the
v
icon on
the lower right-hand portion of the screen (Figure 11A). If the
FEED VTBD
is reprogrammed during a feeding, from
HOLDING
mode,
the pump will restart the feeding and deliver the entire volume of the new VTBD setting.
Figure 11A. ADJUST FEED menu.
VTBD symbol
Figure 11B. Setting the Feed Rate.
Figure 11C. Setting the Feed Volume
to be Delivered.
Continuous Mode Flushing
(Not Available in EZMODE)
For
CONTINUOUS
feeding mode with a feed & flush set loaded, the opening menu will appear as shown in Figure 7B. Press
ADJUST FLUSH
in Figure 7B to program the flushing parameters from the
ADJUST FLUSH
screen, shown in Figure 12A.
Press
FLUSH VOLUME
in Figure 12A to define the volume per flush occurrence. Press the buttons in Figure 12B to program
the flush volume from 10 to 500 in increments of 1 mL. Press
ENTER
(top button) to exit the menu.
Press
FLUSH INTERVAL
in Figure 12A to define the time interval between starts of flushing. Press the buttons in Figure 12C
to program the time interval from 1 to 24 hours in increments of 1 hour. Press
ENTER
(top button) to exit the menu.
The pump will automatically limit flushing settings which exceed pump capabilities to deliver.
Caution:
The Kangaroo Joey enteral feed and flush pump with pole clamp flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use
care when programming the flush volume so that it matches the patient’s need.
Note:
The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as
long as at least one flush had occurred during formula delivery.
Figure 12A. ADJUST FLUSH menu,
continuous mode.
Figure 12B. Setting the
Flush Volume.
Figure 12C. Setting the Flush
Time Interval.