Power Supply/RF Board Replacement
Replacement Pr
ocedur
es
Force FX Electrosurgical Generator C Service Manual
7-29
Install the Power Supply/RF Board Assembly
1.
Position the Power Supply/RF board assembly on the floor of the chassis and install
the four screws that secure the rail to the chassis. Install the screw in the center of the
board (behind the Control board).
2.
Connect the power entry module cable to J6 on the Power Supply/RF board.
3.
Connect the Footswitch board ribbon cable to J5 on the Power Supply/RF board.
4.
Install the front panel assembly.
a.
Position the front panel assembly in front of the chassis and carefully slide it into
the electrical contacts on the Power Supply/RF board.
b.
Install the four screws that secure the front panel to the chassis.
c.
Connect the REM wires to the Power Supply/RF board.
d.
Connect the white wires with the 2-pin nonlocking connector to J15.
e.
Connect the red wires with the 2-pin locking connector to J17.
f.
Connect the power switch cable to J10 on the Power Supply/RF board.
g.
Connect the Display board ribbon cable to the Control board and lock the
connector.
5.
Position the cover above the chassis and slide it down. Install the five screws that
secure the cover to the chassis.
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......