Responding to System Alarms
6-16
Force FX Electrosurgical Generator C Service Manual
174
Software malfunction.
which_errors value outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
180
Internal diagnostics. Main
microcontroller ST
Microelectronics RAM check
failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
181
Internal diagnostics. Feedback
microcontroller ST
Microelectronics RAM check
failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
182
183
184
Internal diagnostics. Main
microcontroller dual-port RAM
check failed
Internal diagnostics. Main
microcontroller Page 0 (zero)
Flash Memory CRC test failed
Internal diagnostics. Main
microcontroller Page F Flash
Memory CRC test failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
185
Internal diagnostics. Feedback
microcontroller Flash Memory
CRC test failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
189
Software malfunction.
display_state value outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
Number
Description
Recommended Action
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......