Using the RS-232 Serial Port
Setup, T
ests, and Adjustment
Force FX Electrosurgical Generator C Service Manual
5-45
Commands for Displaying Data on a Computer
Commands for Changing the Default Coag Mode
No result is displayed unless a character other than D, F, or S is entered Then the
following is displayed:
No Change
Try Again!
To save the new default mode, turn off, then turn on the generator.
Command
Result Displayed on Computer
:PP
Front panel values
:PT
Current real-time clock values
:PSC
Date of last calibration
:PSA
Total time activated in each mode
:PSB
Number of times activated in each mode
:PSK
Number of times output was activated
:PSP
Average power setting used in each mode
:PSF
List of system errors 1 through 10 with time and date stamp
:PSL
List of system errors 11 through 20 with time and date stamp
:PSE#
Number of times specified error has occurred
(# represents an alarm number listed in Chapter 6,
:PG
Generator model and firmware version numbers
Command
Action
:MC <letter>
D
F
S
where <letter> may be any of the following:
sets default coag mode to desiccate
sets default coag mode to fulgurate
sets default coag mode to spray
:MF <value>
0
1
Modifies the selected Fulgurate mode setting, where <value>
represents a selection as follows:
LCF Fulgurate
Fulgurate
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......