Correcting Malfunctions
Troubleshooting
Force FX Electrosurgical Generator C Service Manual
6-5
Generator is on,
but did not
complete the self-
test
An alarm condition exists
Check the display for an alarm number.
Note the number and refer to
Software malfunction
Turn off, then turn on the generator.
Loose or disconnected internal
cables
Check and correct all internal
connections.
Faulty low voltage power
supply
Check the low voltage power supply.
Damaged Control board
connectors
and/or malfunctioning Control
board
Remove the Control board and inspect
the connectors to the Power Supply/RF
board and to the Display board for
damage, poor seating, etc. If the
problem persists, replace the Control
board. For instructions, refer to
Shorts or disconnects on
Power Supply/RF board
Check the Power Supply/RF board for
shorts or disconnects.
Faulty power switch
Replace the power switch. Refer to
Panel Power Switch Replacement
Malfunctioning front panel
components
Replace the front panel assembly. Refer
to
Alarm number
212 appears in
the Cut display
during the self-
test
The battery was removed and/
or replaced, but the generator
was not calibrated
Turn off, then turn on the generator to
clear the number. Calibrate the
generator. For instructions, refer to
Faulty battery on Control
board causing loss of battery-
backed memory
Replace the battery and check the
battery-backed RAM device. Refer to
Correcting Battery-Backed RAM
Malfunctions
Loss of battery-backed RAM
due to faulty component on
the Control board (other than
the battery)
Replace the Control board. Refer to
Continued
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......