Fuse Replacement
7-14
Force FX Electrosurgical Generator C Service Manual
Replacing the Fuse on the Power Supply/RF Board
Equipment required
• Phillips screwdriver
• Fuse puller
1.
Turn off the generator. Disconnect the power cord from the wall receptacle.
2.
Remove the five screws that secure the cover to the chassis. Lift the cover off the
chassis. Set the cover and screws aside for reinstallation.
3.
Locate the fuse at the rear of the Power Supply/RF board, near the power entry
module. Use a fuse puller to remove the blown fuse.
4.
Replace the blown fuse with a 6.3-amp fuse.
5.
Position the cover above the chassis and slide it down. Install the five screws that
secure the cover to the chassis.
Warning
Fire Hazard
For continued protection against fire hazard, replace fuses only with fuses of the
same type and rating as the original fuse.
Electric Shock Hazard
To allow stored energy to dissipate after power is disconnected, wait at
least five minutes before replacing parts.
Precaution
The generator contains electrostatic-sensitive components. When repairing the generator, work at
a static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized generator. Handle circuit boards by their
nonconductive edges. Use an antistatic container for transport of electrostatic-sensitive
components and circuit boards.
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......