Setting Up the Generator
5-6
Force FX Electrosurgical Generator C Service Manual
Connections for Monopolar Surgery
If you plan to use a footswitching monopolar instrument, you must connect a Valleylab
monopolar footswitch. You may also use a footswitch to activate a handswitching
instrument or a CUSA handpiece with CEM nosecone. Refer to the figures on this page
and page 5-7
.
Monopolar connection—footswitch activation and a footswitching or handswitching
instrument using Monopolar 1 Footswitch receptacle and Monopolar 1/CEM Instrument
receptacle
Warning
Electric Shock Hazard
• Do not connect wet accessories to the generator.
• Ensure that all accessories and adapters are correctly connected and that no metal is exposed.
Use only a Valleylab monopolar footswitch with the Force FX Electrosurgical Generator C
generator. Use on an incompatible footswitch may cause unexpected output.
The instrument receptacles on this generator are designed to accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given receptacle.
Doing so will cause simultaneous activation of the instruments.
Valleylab monopolar
footswitch
Patient return electrode
Footswitching or
handswitching
instrument
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......