Correcting Malfunctions
6-10
Force FX Electrosurgical Generator C Service Manual
Abnormal
neuromuscular
stimulation (
stop
surgery
immediately
)
Metal-to-metal sparking
Check all connections to the generator,
patient return electrode, and active
electrodes.
Can occur during coag
Use a lower power setting for the
Fulgurate and Spray modes or select the
Desiccate mode.
Abnormal 50–60 Hz leakage
currents
Inside the generator, carefully inspect for
damage that may cause shorting
between the AC line voltage and
connected patient components.
Situation
Possible Cause
Recommended Action