Calibrating the Generator
Setup, T
ests, and Adjustment
Force FX Electrosurgical Generator C Service Manual
5-41
c.
Stop activation. Verify that the Coag display changed to 120.
d.
Press the coag pedal again and, while activating the generator, press the Coag up
(
) or down (
) arrow buttons until the voltmeter reading is equivalent to
489 ± 6 mA
rms
. Ignore changes to the Coag display.
e.
Stop activation. Verify that the Coag display changed to 10.
6.
To save the ECON calibration values, press the Bipolar up arrow (
) button.
To leave Calibration mode, refer to