Main microcontroller unable
to access GEN_SEM
semaphore
Main microcontroller unable
to access FEEDBACK_SEM
semaphore
Main microcontroller unable
to access ECON_SEM
semaphore
Main microcontroller unable
to access KEY_ACTIVE_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
151
Main microcontroller unable
to access GEN_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
152
Software malfunction.
cal_state value in cs_i_sns_cal(
) outside state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
154
Main microcontroller unable
to access FEEDBACK_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
160
Internal component
malfunction. Dosage error test
failed. ADC values display at
failure
Calibrate the ECON factor. Refer to
.
If the alarm number reappears, refer to
170
Watchdog malfunction.
Correct value not sent from
feedback microcontroller
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
171
172
Watchdog malfunction.
Correct value not sent from
main microcontroller
Watchdog malfunction. Main
microcontroller unable to
access watchdog semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
173
Watchdog malfunction.
Feedback microcontroller
unable to access watchdog
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
Number
Description
Recommended Action
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......