Responding to System Alarms
Troubleshooting
Force FX Electrosurgical Generator C Service Manual
6-19
215
Internal diagnostics. Real-time
clock chip (U1) not compatible
with firmware
1.
Replace the Control board. Refer to
2.
Calibrate the generator. Refer to
If the alarm number reappears, record the
number and call the Covidien Service Center.
216
Internal memory battery
exhausted
1.
Replace the battery. Refer to
2.
Calibrate the generator. Refer to
If the alarm number reappears, record the
number and call the Covidien Service Center.
220
221
Feedback microcontroller
unable to access
KEY_REQ_SEM semaphore
Feedback microcontroller
unable to access GEN_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
222
223
Main microcontroller failed to
copy battery-backed RAM
data to dual-port RAM in time
allotted
Main microcontroller failed to
complete initialization in time
allotted
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
224
Internal diagnostics.
Calibration data checksum
error on feedback
microcontroller (data in error
shown on terminal using serial
port)
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
225
Internal diagnostics. Feedback
microcontroller CPU test failed
1.
Replace the Control board. Refer to
2.
Calibrate the generator. Refer to
If the alarm number reappears, record the
number and call the Covidien Service Center.
Number
Description
Recommended Action
Содержание Force FX
Страница 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Страница 30: ......
Страница 180: ......
Страница 210: ......
Страница 218: ...Generator Assembly 9 4 Force FX Electrosurgical Generator C Service Manual Generator assembly continued...
Страница 222: ...Front Panel Assembly 9 8 Force FX Electrosurgical Generator C Service Manual Front panel assembly continue 7 6 16 17 18...
Страница 246: ......