Calibrating the Generator
5-26
Force FX Electrosurgical Generator C Service Manual
Calibrating the Generator
There are eight calibration steps. During calibration you verify information specific to the
generator, adjust the date, and adjust the clock. You also adjust the REM circuit and
several values, or factors, that ensure the proper operation of the generator.
The calibration steps and the values you can adjust are summarized below. Certain values
are not adjustable but must be verified.
Notice
After completing any calibration step, proceed to the next step to save the values from the
completed calibration step.
Calibration Step and Description
Adjustable?
1.
Generator data
Generator model number
Master microcontroller software version
Feedback microcontroller software version
No (verify value)
No (verify value)
No (verify value)
2.
Calendar
Month
Day of the month
Year
Yes
Yes
Yes
3.
Clock
Hour
Minute
Yes
Yes
4.
REM oscillator frequency and impedance
Yes
5.
Current sense gain factor
Yes
6.
Voltage sense gain factor
Yes
7.
Reactance gain factor
Yes
8.
ECON factor
Yes