Appendix A
Revision 2
A–16
A.4.3
Offset Adjustment
Use the offset to compensate for cable and switch losses. When you initially set
up and test the system, you must force each converter to backup, and adjust the
offset to minimize gain mismatch between the normal and bypassed mode. This
offset will be different for each converter and is stored in a table in the backup
unit.
Do these steps to configure the offset adjustment:
1.
Connect a signal source to the input and a power meter or spectrum
analyzer to the output of the first online converter.
2.
Record the output power.
3.
Force a backup of the first online unit.
4.
Adjust the offset until the output power matches what was recorded in
Step 2.
5.
Restore the online unit, and then repeat these steps for the other online
units.
A.5
Redundancy Systems – Converter Management
See Appendix B. MAINTENANCE
AND
TROUBLESHOOTING for
information about managing the backup and inline converters that
comprise your redundancy system.
Содержание DT-4500 Series
Страница 48: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Specifications Revision 2 2 12 Notes...
Страница 92: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Firmware Update Revision 2 5 16 Notes...
Страница 120: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Front Panel Operation Revision 2 6 28 Notes...
Страница 146: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Ethernet Interface Operation Revision 2 7 26 Notes...
Страница 180: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Serial Interface Operation Revision 2 8 34 Notes...
Страница 204: ...DT 4500 Series C and Ku Band Down Converters MN DT4500 IOM Appendix B Revision 2 B 8 Notes...
Страница 205: ......
Страница 206: ...2114 WEST 7TH STREET TEMPE ARIZONA 85281 USA 480 333 2200 PHONE 480 333 2161 FAX...