71
7.1
Tools >> Device Scan
The Treadmill Software keeps a list of configured devices with which it expects are connected to the CI-
Bus and available for an experiment. This option is used to verify the presence of the configured devices.
Also, it can be used to add/remove devices from the list. This option will open a “Communication Scan”
window:
The following information and controls are available:
“Connected through COM?”: The COM port provided by the CI-Bus Hub where the program found
one or more configured devices upon launch of the program.
Device List Grid: List of all configured or discovered devices on the CI-Bus.
o
Device:
Name of the device.
o
Address#: 1 to 254. The unique CI-Bus address of the device
o
Prod ID#: 0 to 9999. The Product Number of the device (Columbus Instruments specific).
o
Version:
The firmware revision of the device.
o
Status:
“OK” indicates the device has been detected.
“No-Resp” indicates an expected and configured device is absent.
Progress Bar: displays the progress of the “Scan for New Devices” procedure (see below).
Scan Configured Devices : Click this button to verify the presence of all configured devices listed.
Scan for New Devices : Click this button to discover all devices on the CI-Bus.
Their information will display in the device list as they are found.
Exit : Click this button to close the window and return to the main window. If changes have been
made to the list of devices, the user is prompted to save or discard the changes.