Accessible Emission
The table below describes the irradiance in W/cm
2
and appropriate level of eye protection in terms of optical density
(OD) for each product line.
Product
Nominal Output Power
(mW)
Irradiance
(W/cm
2
)*
Eye protection
Requirement**
Cobolt 06-MLD 405 nm
120
47
4
300
117
4
Cobolt 06-MLD 445 nm
100
39
4
Cobolt 06-MLD 473 nm
100
39
4
300
117
4
Cobolt 06-MLD 488 nm
60
23
3
200
78
4
Cobolt 06-MLD 515 nm
80
31
3
Cobolt 06-DPL 532 nm
200
71
3
Cobolt 06-DPL 553 nm
50
18
3
Cobolt 06-DPL 561 nm
100
35
3
Cobolt 06-MLD 633 nm
80
31
3
Cobolt 06-MLD 638 nm
180
70
3
Cobolt 06-MLD 647 nm
130
51
3
Cobolt 06-MLD 660 nm
100
39
3
* Irradiance (W/cm2) = 110% of Nominal Power (W)
Beam Area at bottom tolerance (cm
2
)
** Eye protection (OD) = Log
10
( Max Power (W)
60825-1 Emission Limit : Class 1 (W) ) , rounded up to the next integer.
CAUTION
Always wear the appropriate eye protection for all of the specified emitted wavelengths.
Verify the accessible emission wavelengths and power levels on the warning label before operating.
Fiber Pigtailed Option
All safety recommendations in section 2.1 are also valid for the Cobolt 06-01 series fiber pigtailed laser heads.
Additionally, heat generated from absorption of laser radiation by particles on the fiber end may increase the
probability of ignition hazards in certain environments. Always clean the fiber end before turning on the laser. In
systems where the beam is exposed, fiber end must be mounted < 2 m from the emission warning LED on the key
control box. It is advised to perform a risk assessment for the facility and equipment prior to using the laser. In the
case of integration into a larger system, laser safety compliance must be evaluated in the end product.