OWNERS MANUAL | Cobolt 06-01 Series | D0136-K DECEMBER 2018
201
42 | 52
sdmes
Set digital modulation enable state *
0 = disable
1 = enable
games?
Get analog modulation enable state *
0 = disabled
1 = enabled
sames
Set analog modulation enable state *
0 = disable
1 = enable
gdmes?
Get digital modulation enable state *
0 = disabled
1 = enabled
sames
Set analog modulation enable state *
0 = disable
1 = enable
gom?
Returns the current operating mode
0
–
Off
1
–
Waiting for key
2
–
Continuous
3
–
On/Off Modulation
4
–
Modulation
5
–
Fault
6
–
Aborted
* Analog and digital modulation states are independent. To switch from analog to digital modulation it is necessary
to disable analog and enable digital.
06-MLD Specific Commands
Command
Function
Argument
Returned value
salis
Set analog low impedance (50 Ω) state
0 = disable
1 = enable
galis?
Get analog low impedance (50 Ω)
state
0 = disabled
1 = enabled
06-DPL Specific Commands
Command
Function
Argument
Returned value
smc
Set modulation high current
Float (mA)
gmc?
Get modulation high current
Float (mA)
slth
Set modulation low current
Float (mA)
glth?
Get modulation low current
Float (mA)
stec4t
Set TEC_LD
mod
temperature
Float (°C)
gtec4t?
Get TEC_LD
mod
set temperature
Float (°C)
rtec4t?
Read actual TEC_LD
mod
temperature
Float (°C)