OWNERS MANUAL | Cobolt 06-01 Series | D0136-K DECEMBER 2018
201
47 | 52
06-DPL
Cobolt Monitor™ software expanded to for more detailed monitoring.
TEC Settings
–
shows the running status and the fault status for the laser’s internal thermoelectric coolers (TEC).
TEC
LD
CW
controls the temperature of the platform in CW operation and TEC_LD
mod
controls the platform temperature in
modulation mode.
Laser Operation Mode and Settings
- displays the set laser power. The user can switch between constant power
mode, constant current mode and modulation mode. Likewise, there are boxes to set the constant power level and
constant current level. The output power (measured on an internal photodiode) and the current through the laser
pump diode are both displayed. When in modulation mode it is recommended to use an external power meter, the
power displayed in the software is not actual. See section 6 and 7 for more details on continuous wave and modulation
modes.
Autostart Program
- displays whether the laser is in CDRH or OEM mode and displays the current laser operational
status. There are a
lso buttons to “abort” the autostart sequence or to “restart” the laser after a fault.
NOTE
Specifications are only guaranteed in constant power mode, at 100% of nominal power.
!