Operation and Environmental Specifications
MLD
DPL
Power supply requirements
5 VDC, 3 A
5 VDC, 5 A
Intended use environment
Laboratory (indoor)
Ambient temperature & pointing - 20-50 °C
< 5 µrad /
o
C
Maximum baseplate temperature
50
C
Ambient temperature, operation
10 - 40
º
C
Ambient temperature, storage
-10 -> +60
º
C
Humidity
0-90% RH non-condensing
Ambient Air pressure
950-1050 mbar
Laser Head heat sink thermal impedance at 40
C ambient
< 0.8 K/W
< 0.5 K/W
Maximum heat dissipation of Laser Head
< 12 W
< 20 W
Warm-
up time from complete “off”
< 3 min
Communication protocol
USB and RS 232
USB
Modulation
MLD
DPL
(all)
532 nm
561 nm
553 nm
Digital modulation bandwidth
DC
–
150 MHz
DC
–
50 kHz
DC
–
10 kHz
DC
–
5 kHz
Digital modulation Rise/Fall time
< 2.5 ns
< 6 µs
< 30 µs
< 60 µs
Digital modulation extinction ratio (@10MHz)
> 70 dB
---
Analog modulation bandwidth
DC-2 MHz
DC
–
50 kHz
DC
–
10 kHz
DC
–
5 kHz
Analog modulation extinction ration (@250kHz)
> 70 dB
---
Analog modulation Rise/Fall time
< 300 ns
< 6 µs
< 30 µs
< 60 µs
ON/OFF modulation frequency
DC-500 kHz
---
ON/OFF modulation extinction ratio
inf:1
---
ON/OFF modulation Rise/Fall time
< 300 ns
---