OWNERS MANUAL | Cobolt 06-01 Series | D0136-K DECEMBER 2018
201
41 | 52
Communication commands
The laser is delivered in Auto-start mode (see section 3 for Auto-start sequence description). For system integration
the Auto-start sequence can be disabled and the following commands can be used to control the laser. As long as
power is supplied to the laser the temperature control elements are always operating to reach set-point values and
the laser will be idle waiting for the next command. All arguments are in lower case and separated by a space (ASCII
32).
NOTE
some commands require Auto-start to be disabled but others will work when Auto-start is active.
Command
Function
Argument
Returned value
l?
Get laser ON/OFF state
0 = OFF
1 = ON
@cob1
Restart laser. Forces the laser on
without checking if autostart is
enabled.
hrs?
Get laser head operating hours
Float
ilk?
Get interlock state
0 = OK
1 = interlock open
l1
Laser ON
Requires autostart disabled.
l0
Laser OFF
p?
Get output power set point
Float (W)
P
Set output power
Float (W)
pa?
Read output power
Float (W)
i?
Get drive current
Float (mA)
slc
Set drive current
Float (mA)
cp
Enter constant power mode
ci
Enter constant current mode
f?
Get operating fault
0 - no errors
1
–
temperature error
3 - interlock error
4
–
constant power time
out
cf
Clear fault
gsn?
Get serial number
32-bit unassigned integer
em
Enter modulation mode
!