OWNERS MANUAL | Cobolt 06-01 Series | D0136-K DECEMBER 2018
201
15 | 52
Configuration
CE/CDRH Compliant
The CE/CDRH compliant system is supplied with a key control box, which must be connected, along with a remote
interlock connector. Once power is supplied, laser emission starts when the key is turned from the OFF position to the
ON position. The status of operation can be monitored via LEDs on the key control box. Setting the key to its OFF
position puts the laser in stand-by mode.
The standard CDRH model consists of:
•
Laser head
•
Key control box
•
Keys
•
5 V power supply unit
•
Remote interlock plug (for short circuiting the remote interlock connector)
•
USB communication cable
•
(06-MLD) 15-pin D-SUB male-male cable between the laser head and key control box
•
(06-DPL) 6 pin Molex to 15-pin D-SUB cable between laser head and key control box
OEM
The laser head is supplied
without
the key control box. Connecting a 5 VDC power supply to the laser head initiates
an automatic start-up sequence. If the remote interlock is connected, laser emission will start automatically as soon
as power is supplied and internal temperatures are stabilized.
The OEM model consists of:
•
Laser head
•
5 V power supply unit
•
USB communication cable
•
Remote interlock plug (for short circuiting the remote interlock connector)