Section 12: Troubleshooting / General Problems with the Analyzer
1000HN User’s Guide / page 60
3.
If the microprocessor assembly feels warm and the display has a blue back-
ground, but is otherwise blank, adjust the Display Intensity control at the back of
the analzyer by turning it clockwise. If the display is still blank, call Cirris to
replace the microprocessor assembly.
4.
If the microprocessor assembly is warm, but the display is totally blank, or there
is no blue background showing in the display, call Cirris to replace the
microprocessor assembly.
Problem
: Display shows a darkened row, or gibberish characters.
Possible Causes
:
• Display Intensity control is not adjusted properly.
• Scanner assembly is bad.
• Microprocessor assembly is bad.
Things to do
:
1.
If both rows of the display are dark, try adjusting the Display Intensity control at
the back of the analyzer by turning it counterclockwise. If this does not solve the
problem, go to step 2.
2.
Turn off the analyzer, disconnect the microprocessor from the rest of the system,
then turn on the microprocessor. The analyzer should prompt
10 V LO
. If it
does, the analog board or one of the scanner boards is bad. Call Cirris to replace
it.
3.
If the problem is not solved after the microprocessor is disconnected, the micro-
processor assembly is probably bad. Call Cirris to replace it.
Problem
: When a cable is disconnected, the display still prompts
Please Verify
.
Possible Causes:
• A cable adapter is bad, or adapting cables are bad.
• The scanner assembly is defective.
Things to do
:
1.
Remove the cable adapters from the analyzer while the analyzer is still on. If the
analyzer still prompts
Please Verify
, the scanner is probably defective. Call
Cirris to replace it.
2.
If the analyzer prompts
Ready To Test
after step one is done, you have
unwanted connections in your cable adapters or adapting cables.
Display Intensity control
Содержание Signature 1000HN
Страница 1: ...Signature 1000HN Cable Analyzer User s Guide Version 4 0 Major Revision 1 December 1999...
Страница 4: ......
Страница 6: ......
Страница 8: ...Introduction to the Cirris 1000HN What your order should contain 1000HN User s Guide page 8...
Страница 16: ...Section 1 Work With the Hardware Changing the company name 1000HN User s Guide page 16...
Страница 22: ...Section 3 What the Option Settings Mean Auto Print 1000HN User s Guide page 22...
Страница 48: ...Section 10 Select a Test Procedure Tips on selecting Insulation Resistance settings 1000HN User s Guide page 48...
Страница 66: ...Section 12 Troubleshooting Problems with Printing 1000HN User s Guide page 66...
Страница 68: ...Section 13 Specifications Warranty 1000HN User s Guide page 68...
Страница 70: ...Section 14 Statement of Warranty 1000HN User s Guide page 70...
Страница 74: ...Section 15 Glossary short 1000H User s Guide page 74...
Страница 76: ...Section 16 Blank Forms to Photocopy 1000HN User s Guide page 76...
Страница 77: ...Master Parts List Cable Part Number Description Signature Copyright 1999 by Cirris Systems Corporation...
Страница 78: ......
Страница 80: ......
Страница 85: ...W Warranty 69...
Страница 86: ......