1000HN User’s Guide / page 19
Overview
The Cirris 1000HN has eleven test option settings which you set to meet your testing
requirements. In this section, we’ll tell you what each of the settings means.
Create Test From
Gives you the option to create a new test setup from a new Sample Cable, your last
test setup, or by learning a complex cable that contains resistors and/or diodes in addi-
tion to ordinary wire connections.
SAMPLE CABLE
: Use this setting to learn standard simple cables (cables that
contain only wires and connectors).
LAST TEST SETUP
: Use this setting when you need to use the same test setup each
time you turn on the analyzer. Useful when testing large batches of the same type
cable. Will help protect your test setup from a power failure, or if you accidentally
turn on the analyzer with an untested cable connected.
COMPLEX ASSEMBLY
: Use this setting to learn a cable that contains resistors,
diodes, or divider networks in addition to ordinary wires and connectors. When you
use this setting, the Connection Resistance refers to the resistance level that definitely
separates wires from resistors. The analyzer sets this to 5
Ω.
You may change the
setting to meet your needs. Using the Connection Resistance CALC mode may help
you determine a good setting.
Connection
Resistance
Specifies the maximum resistance a connection can have and still be considered
good. For example, if you set the Connection Resistance to 10
Ω
, the analyzer will
consider all connections with resistances less than 10
Ω
as good, and those with resis-
tances greater than 10
Ω
as bad.
Range
:.5
Ω, 1Ω, 2Ω, 5Ω, 10Ω, 20Ω,50Ω, 100Ω, 200Ω, 500Ω, 1ΚΩ,
AUTO, CALC.
AUTO mode
: When you set the Connection Resistance to AUTO, the 1000HN
measures the resistance of each connection in your Sample Cable, then automatically
sets the Connection Resistance threshold to a value 20% higher than the largest
resistance measured in the Sample Cable. The 20% margin of error is increased for
resistances less than 1
Ω.
CALC mode
: Setting the Connection Resistance to
CALC
causes the 1000HN to
measure the resistance of the connections in the Sample Cable in the same way you
would measure them with an ohmmeter. In this mode, the analyzer can only measure
resistances up to 6K
Ω
. After the analyzer calculates the resistance of the Sample
Cable, it prompts
RESISTANCE READY
. Press the
Display/Print
switch to display
all resistances measured below 6Kohms.
Hipot Voltage
Range
: Off, 50V, 100V, 200V, 300V, 400V, 500V, 630V. Specifies what voltage will
be applied during hipot testing, between pins that should not be connected. This deter-
Section 3:
What the
Option
Settings Mean
Содержание Signature 1000HN
Страница 1: ...Signature 1000HN Cable Analyzer User s Guide Version 4 0 Major Revision 1 December 1999...
Страница 4: ......
Страница 6: ......
Страница 8: ...Introduction to the Cirris 1000HN What your order should contain 1000HN User s Guide page 8...
Страница 16: ...Section 1 Work With the Hardware Changing the company name 1000HN User s Guide page 16...
Страница 22: ...Section 3 What the Option Settings Mean Auto Print 1000HN User s Guide page 22...
Страница 48: ...Section 10 Select a Test Procedure Tips on selecting Insulation Resistance settings 1000HN User s Guide page 48...
Страница 66: ...Section 12 Troubleshooting Problems with Printing 1000HN User s Guide page 66...
Страница 68: ...Section 13 Specifications Warranty 1000HN User s Guide page 68...
Страница 70: ...Section 14 Statement of Warranty 1000HN User s Guide page 70...
Страница 74: ...Section 15 Glossary short 1000H User s Guide page 74...
Страница 76: ...Section 16 Blank Forms to Photocopy 1000HN User s Guide page 76...
Страница 77: ...Master Parts List Cable Part Number Description Signature Copyright 1999 by Cirris Systems Corporation...
Страница 78: ......
Страница 80: ......
Страница 85: ...W Warranty 69...
Страница 86: ......